XACT® 640 MULTI-METALS MONITOR
XACT® 640 MULTI-METALS MONITOR
Cooper Environmental Service¥100
Cooper Environmental’s Xact® 640 is the first Multi-Metals Continuous Emissions Monitor to be proven successful for accurately measuring elemental concentrations in smoke stacks by capturing and measuring both particulate and vapor phase metals.
Out of stock
- Description
- Specification
- Application and features
Product Description
Cooper Environmental’s Xact® 640 is the first Multi-Metals Continuous Emissions Monitor to be proven successful for accurately measuring elemental concentrations in smoke stacks by capturing and measuring both particulate and vapor phase metals.
The Xact® 640 system uses reel-to-reel filter tape sampling and nondestructive X-ray fluorescence (XRF) analysis to monitor stack hazardous air pollutant (HAP) metal emissions. An isokinetic sub-sample of stack gas goes into the stilling chamber and then drawn through a chemically reactive filter tape. Vapor phase metals, including mercury, are deposited along with particulate-phase metals on the filter tape.
In 2007, the US EPA recognized the Xact® CEMS as an outstanding achievement in innovative clean air technology with a clean air excellence Award. In addition, the EPA approved the Xact® 640 as an alternative method for periodic Method 29 testing, feed stream analysis, and monitoring emissions during plant operation.
Measurement method:
– EPA method IO3.3: Determination of metals in ambient PM using XRF
Key applicable elements:
-Sb, As, Ba, Cd, Ca Cr, Co, Cu, Fe, Pb, Hg, Mn, Ni, Se, Ag, Sn, Ti, Tl, V, Zn, and more available.
–
Measurement range:
-Demonstrated up to 1963ug/dscm
Detection limit:
-Metal and sample time dependent; refer to the minimum detection limit(MDL) data.
Calibration stability check frequency:
-Automatically with each sample analyzed
Sampling and analysis times:
-Every 15,30, 60 or 120 minutes
Estimated recalibration frequency:
-Annually, when manufacturer’s operation recommendations followed
Sample flowrate:
– – Application specific
Linearity:
-Correlation coefficient > 0.98
Size and weight:
– 19” x 24” x 19”
– 180 lbs
– 483 mm rack-mountable components
Required operating environment:
-Lab environment with temperature controlled to 20±3°C (68°F)/>
– 220 VAC/60Hz 20A
Power requirement:
-120 VAC/60Hz @ 2-20 amp
or:
-220 VAC/60Hz @ 20 amp with an optional power converter
Output:
– RS232, Modbus protocol
Options:
– Change or add elements
– Enclosure
– Remote control
– Remote polling
Minimum Detection Limits (ppb)
Sampling Time (min.)
Element Atomic Number 15 30 60 120
Cr 24 0.14 0.05 0.018 0.006
Mn 25 0.14 0.05 0.018 0.006
Fe 26 0.38 0.13 0.048 0.017
Co 27 0.16 0.06 0.020 0.007
Ni 28 0.11 0.04 0.014 0.005
Cu 29 0.13 0.05 0.017 0.006
Zn 30 0.12 0.04 0.014 0.005
Ga 31 0.05 0.02 0.007 0.002
Ge 32 0.06 0.02 0.008 0.003
As 33 0.06 0.02 0.007 0.003
Se 34 0.07 0.02 0.009 0.003
Ag 47 2.17 0.77 0.271 0.096
Cd 48 2.88 1.02 0.360 0.127
Sn 50 3.74 1.32 0.467 0.165
Sb 51 0.33 0.12 0.042 0.015
Ba 56 0.47 0.17 0.059 0.021
Hg 80 0.09 0.03 0.012 0.004
Tl 81 0.09 0.03 0.012 0.004
Pb 82 0.11 0.04 0.014 0.005
Bi 83 0.12 0.04 0.015 0.005
• Automatic quality assurance, alarms and control features
• Gas phase calibration not required
• Identification and measurement of 23 elements simultaneously (refer to the periodic table on the 640 Spec Sheet below)
• Internal calibration check incorporated with every sample analyzed
• Proven RTR XRF technology demonstrated on the ocean floor, Mars, and in thousands of beta attenuation monitors
• Daily automatic upscale, blank, and flow checks
• Recognized by the EPA as an innovative clean air technology (Clean Air Excellence Award, 2007)
• Sampling, analysis, and near-real-time reporting (every 15, 30, 60 and 120 minutes)